Interferometric system based on swept source-optical coherence tomography scheme applied to the measurement of distances of industrial interest
Date
2015-12-11
Journal Title
Journal ISSN
Volume Title
Publisher
Abstract
We used an interferometric technique based on typical optical coherence tomography (OCT)
schemes for measuring distances of industrial interest. The system employed as a light source a tunable erbium-doped fiber laser of ∼20-pm bandwidth with a tuning range between 1520 and 1570 nm. It has a sufficiently long coherence length to enable long depth range imaging. A set of fiber Bragg gratings was used as a self-calibration method, which has the advantage of being a passive system that requires no additional electronic devices. The proposed configuration and the coherence length of the laser enlarge the range of máximum distances that can be measured with the common OCT configuration, maintaining a good axial resolution.
A measuring range slightly >17 cm was d0.etermined. The system performance was evaluated by studying the repeatability and axial resolution of the results when the same optical path difference was measured.
Additionally, the thickness of a semitransparent medium was also measured.
Description
Keywords
UTN, Facultad Regional Delta, Optical coherence tomography, Metrological instrumentation, Distance measurement.
Citation
Interferometric system based on swept source-optical coherence tomography scheme applied to the measurement of distances of industrial interest.
Collections
Endorsement
Review
Supplemented By
Referenced By
Creative Commons license
Except where otherwised noted, this item's license is described as info:eu-repo/semantics/openAccess